Product Overview
The MPP500 Multi-Parameter Water Quality Analyzer is an integrated online monitoring system designed for drinking water and municipal water quality applications. It combines multiple sensors into one wall-mounted analyzer to continuously monitor key water quality parameters while reducing installation complexity and maintenance costs. Ideal for water treatment plants, distribution networks, secondary water supply systems, and public water facilities.
Key Features
Technical Specifications
System
- Power Supply: (110±11) VAC, (50±1) Hz
- Power Consumption: ≤30 W
- Cabinet Size: 19.7 × 15.7 × 7.9 in (500 × 400 × 200 mm)
- Weight: Approx. 26.5 lb (12 kg)
- Storage Temperature: 39.2–122°F (4–50°C)
- Operating Temperature: 32–122°F (0–50°C)
- Operating Humidity: ≤95% RH (Non-condensing)
- Inlet Water Flow: 0.13–0.26 GPM (0.03–0.06 m³/h)
- Inlet Pressure: <43.5 psi (<0.3 MPa)
- Display: Color Touch Screen
- Communication Interface: RS-485 Modbus RTU + Wireless Interface
Turbidity
- Range:0-1NTU / 0-20NTU / 0-100NTU / 0-2000NTU
- Resolution:0-1NTU / 0-20NTU / 0-100NTU:0.001NTU0-2000NTU:0.01NTU
- Detection limit:0.005 NTU
- Zero drift:≤1.5%
- Indication stability:≤1.5%
- Accuracy:2% or ±0.02NTU; 10% or 0.5NTU (0-2000NTU)
- Repeatability:≤1%
- Response time:T90≤120s
- Recommended maintenance period:3-12 months (depending on the water quality on site)
Residual chlorine/chlorine dioxide/ozone
- Range:0-5mg/L / 0-20mg/L
- Resolution:0.01mg/L
- Detection limit:0.05mg/L
- Accuracy:±0.05mg/L or ±5% (DPD comparison error ±10%)
- Response time:≤120 seconds
- Recommended maintenance period:1-3 months or weekly calibration, 3-6 months to replace consumables
pH /ORP(optional)
- Range:0-14 pH, ±2000 mV (ORP)
- Resolution:0.01 pH, ±1 mV (ORP)
- Accuracy:±0.1 pH, ±20 mV (ORP) or ±2%
- Repeatability:±0.1 pH, ±10 mV (ORP)
- Response time:≤60 seconds
- Recommended maintenance period:1-3 months
Temperature
- Range:32–122°F (0–50°C)
- Resolution:0.18°F (0.1°C)
- Accuracy:±0.9°F(±0.5°C)
- Repeatability:≤0.9°F (≤0.5°C)
- Response Time:≤25 s
- Recommended Maintenance Interval:12 Months
Conductivity (Optional)
- Range:0-20000uS/cm; 0-20uS/cm(Pure water sensor)
- Accuracy:±0.8%F.S; ±3%F.S(Pure water sensor)
- Resolution:0.01uS/cm
- Repeatability:≤0.4%FS
- Response time:≤30 seconds
- Recommended maintenance period:3-6 months
Applications