Introduction

The Multi-parameter analyzer can be widely used in urban or rural water supply plants, tap water pipeline networks, tap water secondary water supply, user taps, indoor swimming pools, Online monitoring of water quality such as large-scale water purification equipment and direct drinking water is an indispensable online analysis equipment in the fields of water plant production process control, water conservancy and water management, and sanitation supervision.

Details

System

  • working power (220±22)VAC, (50±1)Hz
  • Power 30W
  • Cabinet size 800mm*506mm*180mm (standard version)
  • Weight about 15Kg
  • Storage temperature 4℃~+50℃
  • Working temperature 4℃~+50℃ / -25℃~+50℃(optional temperature control heating antifreeze module)
  • Working humidity ≤95%RH (no condensation)
  • Inlet flow 500-1000 mL/min
  • Inlet pressure < 3kg/cm³
  • Communication interface RS485 Modbus RTU communication protocol + wireless interface

Turbidity

  • Range 0-1NTU / 0-20NTU / 0-100NTU / 0-4000NTU
  • Resolution 0-1NTU / 0-20NTU / 0-100NTU:0.001NTU 0-4000NTU:0.01NTU
  • Lower detection limit 0.02 NTU; 0.1NTU (0-4000NTU)
  • Zero drift ≤1.5%
  • Indication stability ≤1.5%
  • Accuracy 2% or ±0.02NTU; 5% or 0.5NTU (0-4000NTU)
  • Repeatability ≤3%
  • Response time T90≤120s
  • Recommended maintenance period 3-12 months (depending on the water quality on site)

Residual chlorine/chlorine dioxide

  • Range 0-5mg/L / 0-20mg/L
  • Resolution 0.01mg/L
  • Lower detection limit 0.05mg/L
  • Accuracy ±0.05mg/L or ±5% (DPD comparison error ±10%)
  • Response time ≤120 seconds
  • Recommended maintenance period 1-3 months or weekly calibration, 3-6 months to replace consumables

PH /ORP(optional)

  • Range 0-14 pH, ±2000 mV (ORP)
  • Resolution 0.01 pH, ±1 mV (ORP)
  • Accuracy ±0.1 pH, ±20 mV (ORP) or ±2%
  • Repeatability ±0.1 pH, ±10 mV (ORP)
  • Response time ≤60 seconds
  • Recommended maintenance period 1-3 months

Temperature

  • Range 0℃ - 50℃
  • Resolution 0.1℃
  • Accuracy ±0.5℃
  • Repeatability ≤0.5℃
  • Response time ≤25 seconds
  • Recommended maintenance period 12 months

Conductivity (Optional)

  • Range 0-2000uS/cm 
  • Accuracy ±1.5%FS
  • Repeatability ≤0.5%FS
  • Response time ≤30 seconds
  • Recommended maintenance period 3-6 months

Dissolved oxygen (Optional)

  • Range 0-20mg/L
  • Accuracy ±0.3mg/L
  • Repeatability ≤±1.5%
  • Response time ≤30 seconds
  • Recommended maintenance period 1-3 months

Expansion port

  • Port type RS485, 4-20mA