This is an all-inclusive reference price (product, international shipping, and import duties), based on a few selected configurations. Actual pricing may vary by size, options, and order quantity. Minimum order may apply for low-value items, while better pricing may be available for larger requirements. For a confirmed quote or to place an order, please contact us.
Measurement parameter
View full details

Introduction


The MPP500 Multi-Parameter Water Quality Analyzer is an integrated online monitoring system designed for drinking water and municipal water quality applications. It combines multiple sensors into one wall-mounted analyzer to continuously monitor key water quality parameters while reducing installation complexity and maintenance costs. Ideal for water treatment plants, distribution networks, secondary water supply systems, and public water facilities.

Technical Specifications

System

  • Power Supply: (110±11) VAC, (50±1) Hz
  • Power Consumption: ≤30 W
  • Cabinet Size: 19.7 × 15.7 × 7.9 in (500 × 400 × 200 mm)
  • Weight: Approx. 26.5 lb (12 kg)
  • Storage Temperature: 39.2–122°F (4–50°C)
  • Operating Temperature: 32–122°F (0–50°C)
  • Operating Humidity: ≤95% RH (Non-condensing)
  • Inlet Water Flow: 0.13–0.26 GPM (0.03–0.06 m³/h)
  • Inlet Pressure: <43.5 psi (<0.3 MPa)
  • Display: Color Touch Screen
  • Communication Interface: RS-485 Modbus RTU + Wireless Interface

Turbidity

  • Range:0-1NTU / 0-20NTU / 0-100NTU / 0-2000NTU
  • Resolution:0-1NTU / 0-20NTU / 0-100NTU:0.001NTU0-2000NTU:0.01NTU
  • Detection limit:0.005 NTU
  • Zero drift:≤1.5%
  • Indication stability:≤1.5%
  • Accuracy:2% or ±0.02NTU; 10% or 0.5NTU (0-2000NTU)
  • Repeatability:≤1%
  • Response time:T90≤120s
  • Recommended maintenance period:3-12 months (depending on the water quality on site)

Residual chlorine/chlorine dioxide/ozone

  • Range:0-5mg/L / 0-20mg/L
  • Resolution:0.01mg/L
  • Detection limit:0.05mg/L
  • Accuracy:±0.05mg/L or ±5% (DPD comparison error ±10%)
  • Response time:≤120 seconds
  • Recommended maintenance period:1-3 months or weekly calibration, 3-6 months to replace consumables

pH /ORP(optional)

  • Range:0-14 pH, ±2000 mV (ORP)
  • Resolution:0.01 pH, ±1 mV (ORP)
  • Accuracy:±0.1 pH, ±20 mV (ORP) or ±2%
  • Repeatability:±0.1 pH, ±10 mV (ORP)
  • Response time:≤60 seconds
  • Recommended maintenance period:1-3 months

Temperature

  • Range:32–122°F (0–50°C)
  • Resolution:0.18°F (0.1°C)
  • Accuracy:±0.9°F(±0.5°C)
  • Repeatability:≤0.9°F (≤0.5°C)
  • Response Time:≤25 s
  • Recommended Maintenance Interval:12 Months

Conductivity (Optional)

  • Range:0-20000uS/cm;  0-20uS/cm(Pure water sensor)
  • Accuracy:±0.8%F.S; ±3%F.S(Pure water sensor)
  • Resolution:0.01uS/cm
  • Repeatability:≤0.4%FS
  • Response time:≤30 seconds
  • Recommended maintenance period:3-6 months

Key Features

  • Monitor up to five critical water quality parameters simultaneously through a single integrated platform for efficient online monitoring.
  • Integrated cabinet with centralized water inlet and outlet minimizes installation space while protecting the system from flooding and moisture.
  • Designed for continuous monitoring of drinking water with stable performance for low turbidity and trace disinfectant measurement.
  • Supports automatic drainage and remote adjustment functions, helping reduce routine maintenance frequency and operating costs.
  • Supports RS485 Modbus RTU communication with optional wireless data transmission for seamless SCADA integration.
  • Available in standard and customized versions with flexible parameter combinations, cabinet appearance, branding, and interface options.

Applications

FAQ

What parameters can the MPP500 monitor?

The MPP500 supports configurable combinations of pH, turbidity, residual chlorine, chlorine dioxide, ozone, conductivity/TDS, temperature, ORP, and other optional parameters depending on your application.

Can I customize the monitored parameters?

Yes. The analyzer is available in standard and customized configurations, allowing different parameter combinations and customized cabinet branding.

Is the MPP500 suitable for outdoor installation?

Yes. The wall-mounted IP65 enclosure is designed for outdoor environments, and an optional anti-freeze heating module is available for cold climates.

How often does the analyzer require maintenance?

Maintenance intervals depend on the monitored sensors and water quality. Most sensors require routine inspection every 1–6 months, while turbidity sensors may require maintenance every 3–12 months.

What warranty do you provide?

Please refer to BayMeters Warranty Policy or contact our sales team for warranty details.

How long does shipping take?

Products in the US warehouse typically ship within 3–5 business days

Contact form